Advanced Chemical Sensors, Inc.
Technical Information Sheet
Acrolein Vapor Monitor
Catalog No. OV-055
ACS Vapor Monitor can be worn as a badge near the breathing zone
to measure personal exposure, or placed in a room to measure area
concentration.
Occupational safety standards are based on personal exposure
monitoring of each employee who may be exposed to the vapor as a
regular part of their job. Periodic monitoring of personnel and room
concentrations is recommended
| Exposure Limits: |
ACGIH |
OSHA |
NIOSH |
| TWA (ppm) |
N/A |
0.1 |
0.1 |
| STEL/CEIL (ppm) |
0.1 |
N/A |
0.3 |
| Sampling Medium: |
Chemically Activated Silica Gel |
| Method of Analysis: |
High Performance Liquid Chromatography (HPLC) |
| Recommended Sampling Time: |
15 minutes to 8 hours |
| Minimum Level of Quantitation: |
0.01 ppm for 8 hours
0.3 ppm for 15 minutes; 0.1 ppm with field blank |
| Interferences: |
No interferences are known. |
| Effect of Temperature: |
Less than 10% for each 10 °C variation from 24 °C |
| Effect of Humidity: |
No effects detected at 30% RH to 92% RH. |
| Overall System Accuracy: |
18% at 0.1 ppm for 8 hours |
| Storage Conditions: |
Maximum shelf life 6 months. |
| Recommended Holding Time: |
One week prior to analysis. |
|