[an error occurred while processing this directive]
Accurate, Cost-Effective Chemical Analysis and Monitoring

Advanced Chemical Sensors, Inc.
Technical Information Sheet

Propargyl Alcohol Vapor Monitor
Catalog No. OV-87

ACS Vapor Monitor can be worn as a badge near the breathing zone to measure personal exposure, or placed in a room to measure area concentration.

Occupational safety standards are based on personal exposure monitoring of each employee who may be exposed to the vapor as a regular part of their job. Periodic monitoring of personnel and room concentrations is recommended


Exposure Limits: ACGIH OSHA NIOSH
TWA (ppm) 1 N/A 1
STEL/CEIL (ppm) N/A N/A N/A

Sampling Medium: Chemically Activated Carbon Molecular Sieve
Method of Analysis: Gas Chromatography
Reference Method: OSHA Method 97
Recommended Sampling Time: 15 minutes to 8 hours
Minimum Level of Quantitation: 0.01 ppm for 8 hours
Interferences: No interferences are known.
Effect of Temperature: Less than 10% for each 10 °C variation from 24 °C
Effect of Humidity: No effects detected at 30% RH to 92% RH.
Overall System Accuracy: 16% at 1 ppm for 8 hours
Storage Conditions: Maximum shelf life one year.
Recommended Holding Time: One week prior to analysis.