Technical Information Sheet
o-Phthalaldehyde Vapor Monitor
Catalog No. OP-10
ACS Vapor Monitor can be worn as a badge near the breathing zone
to measure personal exposure, or placed in a room to measure area
concentration.
Occupational safety standards are based on personal exposure
monitoring of each employee who may be exposed to the vapor as a
regular part of their job. Periodic monitoring of personnel and room
concentrations is recommended
| Exposure Limits: |
ACGIH |
OSHA |
NIOSH |
| TWA (ppm) |
N/A |
N/A |
N/A |
| STEL/CEIL (ppm) |
N/A |
N/A |
N/A |
| Sampling Medium: |
Chemically Activated Adsorbent to form a derivative of o-phthalaldehyde |
| Method of Analysis: |
Liquid Chromatography |
| Recommended Sampling Time: |
15 minutes to 8 hours |
| Minimum Level of Quantitation: |
0.0002 ppm for 8 hours
0.06 pm for 15 minutes
|
| Interferences: |
No interferences are known.
Alcohols, other aldehydes and ketones will not interfere.
|
| Effect of Temperature: |
Less than 10% for each 10 °C variation from 24 °C |
| Effect of Humidity: |
No effects detected at 30% RH to 92% RH. |
| Overall System Accuracy: |
9.2% at 1.0 ppm for 8 hours
13% at 1.0 ppm for 15 minutes
|
| Storage Conditions: |
Maximum shelf life six months. |
| Recommended Holding Time: |
Two weeks prior to analysis. |
|